CAT NO: DIDACTICNLE-Export-184002
Steri High Pressure High Speed Horizontal Cylindrical/Rectangular Regulated Medical Waste Sterilizer
Purpose :–
Equipment
is designed for sterilization of regulated biomedical waste generated
from hospital for destruction of bacteria and related organism (Spores).
Design :–
The sterilizer equipment to treat medical waste is designed for following parameters.
For
gravity flow autoclave : The temperature of not less than 121°C and
pressure of not less than 1.02 K.G./CM2 and autoclave residence time not
less than 60 minutes.
The temperature of not less 135°C and
pressure of not less than 2.1 K.G./CM2 and autoclave residence time not
less than 45 minute.
The temperature of not less than 149°C and
pressure of not less than 3.5 K.G./CM2 and autoclave residence time not
less than 30 minute.
Available in following Sizes :–
Cylindrical Model :
400 (Dia) × 600 mm (Depth)
500 (Dia) × 1200 mm (Depth)
Rectangular Model :
600 (H) × 600 (W) × 1200 mm (D)
900 (H) × 600 (W) × 1500 mm (D)
Other custom built sizes are also available.