CAT NO: DIDACTICNEL0448
Pharmaceutical Stability Test Chambers.
Pharmaceutical Stability Test Chambers Technical data: The cabinets should meet the requirements of ICH Guidelines Q1A (Stability Testing) and Q1B (Photostability Testing).
Temperature range: +10 to +60°C Humidity range: 20 to 90%RH Temperature and Humidity accuracy ± 0.3°C / ±2.5%RH.
Temperature and Humidity uniformity ± 0.5°C / ±3.0%RH. PID Cotrolled with LCD display. Humidity sensor.
Built in Digital Recorder. Balanced Temperature and Humidity Control system (B.T.H.C).
Running Hours of operation counter. Safety Device: ELB, Refrigerator Overload Relay, Overheat Protector, Water Empty