CAT NO: DIDACTICNEL0069
Infinity Refractive Index Detector.
infinity Refractive index Detector
Includes 8uL flow cell and LAN interface and twisted pair cable.
Detection type: Deflection method
Short-term noise: ± 2.5 x 10-9 RIU
Drift: < 200 x 10-9 RIU/h
Refractive index range: 1.00-1.75, calibrated
Flow cell: 8 µL,5 bar pressure maximum
Temperature control: Ambient + 50C to 550C
pH range: 2.3-9.5
Time programmable: Polarity, peak width
Zero adjust: Automatic zero
Valves: Automatic purge and automatic solvent recycle
Data rate: UP to 37 Hz
Analog output: Recorder/integrator: 100mV or 1V, with offset adjustment,
RIV range selectable.
Communications: Local Area Network (LAN), Control Area Network (CAN),
RS-232, APG remote, remote ready, start and shutdown signa
Safety and maintenance: Extensive diagnostics, error detection and
display leak detection, safe leak handling, leak output signal for
shutdown of pumping system, low voltage in major maintenance areas.
GLP features: Early maintenance feedback (EMF) for continuous tracking
of instrument usage with user-settable limited (purge interval) and
feedback messages. Electronic records of maintenance and errors,
automated OQ/PV procedures.