CAT NO: DIDACTICNLE-Export-009069
1. Provided with refrigration system to
control temperature range from 10C to 60C ± 2C and humidity 55% to 95% ±
5%.
2. Temperature and Humidity can be
controlled through Microprocessor Based Digital Temp. & Humidity
Indicator cum controller.
3. Double wall,
inner made of Stainless Steel, outer made of thick best quality Mild
Steel with powder coated.
4. Insulation between inner and outer wall is done with glass wool to prevent any heat loss.
5.
Inner chamber is provided with 3Nos. perforated Stainless Steel shelves
with ribs to adjust the shelves at desirable height.
6. Provided with Air Circulation fan in the chamber for uniformity of temperature.
7. Timer 1minute to 99hrs.
8.
Provided with double door,inner made of thick transparent full view
glass and outer door is double walled metallic properly insulated with
glass wool and also provided with best quality handle,gasket
etc.
9. With water level indicator.
10. Whole unit is mounted on Castor wheels.
11. Interior Chamber with fluorescence tubes.
12. Shape & Size can be vertical in size with panel attached on top or side panel(as per client requirement).
13. Leakproof positive locking arrangement will be there.
14. CFC free ECO friendly compressor.
15. To work on 220/230volts AC
Temperature Range : 20°C to 60°C
Accuracy : ± 0.2°C
Humidity Range :40% RH to 95% RH
Accuracy : ±2% RH
Temperature Uniformity : ±1°C
Humidity Uniformity : ± 3% RH
Temperature & Humidity Suitable for 25 °C& 60% RH, 40 °C & 75% RH, 30 °C & 65% RH
Specially
designed chambers to carry out stability studies at low humidity
conditions such as 25°C & 40% RH, 40°C & 25% RH NMT, 30°C &
35% RH
Humidity System
Humidity is by steam injection.
Humidity system consist of boiler tank, which is fitted with two nos. of
boiler heaters, one heater works as a stand by and automatic switch
over to stand by heater if regular fails and give alarm.A water
reservoir tank is connected to the boiler tank. DM water is preferred. A
magnetic float switch provided for safety for boiler low level to cut
off the its supply.
Alarms
Audio and visual alarms are generated for temperature / humidity variation, utility failure.
Safety
The unit is provided with safety devices for temperature, and humidity over shoot in case of malfunction..